Turbidity Sensor Calibration Procedure at Instronline
Friday, September 14, 2018 IST
The turbidity method is based upon a comparison of intensity of light scattered by a sample under defined conditions with the intensity of light scattered by a standard reference suspension. A turbidimeter is a nephelometer with a visible light source for illuminating the sample and one or more photo-electric detectors placed ninety degrees to the path of the light source. Note: the below calibration procedure is for a turbidimeter which the sample is placed into a cuvette.
Some instruments will only accept one standard. For those instruments, the second, third, etc., standards will serve as check points.
Calibration Procedure:
1. Allow the calibration standards to equilibrate at the ambient temperature. The use of commercially available polymer primary standards (AMCO-AEPA-l) is preferred; however, the standards can be prepared using Formazin (read the warning on the label before use) according to the EPA analytical Method 180.1. Other standards may be used if they can be shown that they are equivalent to the previously mentioned standards.
2. If the standard cuvette is not sealed, rinse a cuvette with deionized water. Shake the cuvette to remove as much water as possible. Do not wipe dry the inside of the cuvette because lint from the wipe may remain in the cuvette. Add the standard to the cuvette.
3. Before performing the calibration procedure, make sure the cuvettes are not scratched and the outside surfaces are dry and free from fingerprints and dust. If the cuvette is scratched or dirty, discard or clean the cuvette respectively. Note: some . manufacturers require the cuvette to be orientated in the instrument in a particular direction for accurate reading.
4. Select a low value standard such as a zero or 0.02 NTU and calibrate according to manufacturer’s instructions. Note: a zero standard (approximately 0 NTU) can be prepared by passing distilled water through a 0.45 micron pore size membrane filter.
5. Select a high standard and calibrate according to manufacturer’s instructions or verify the calibration if instrument will not accept a second standard. In verifying, the instrument should read the standard value to within the specifications of the instrument. If the instrument has range of scales, check each range that will be used during the sampling event with a standard that falls within that range.
6. Record the calibration information on calibration log sheet.
Related Topics
Related News & Articles
Fluid flow meters are divided into two functional groups – One measures quantity (Positive Displacement); the other measures rate of flow (Inferential.) All...
Recently posted . 774 views . 3 min read
Differential pressure sensors are common in the process industry and cover a variety of applications. To understand what a differential pressure sensor is, it bec...
Recently posted . 815 views . 3 min read
The 2018 iPhones might be still in the rumor mill, but information about the 2019 devices is already leaking. According to sources from Taiwan, at least one of th...
Recently posted . 782 views . 2 min read
The Lenovo wearable is now available from Flipkart.
Recently posted . 952 views . 0 min read
The new sensor could greatly enhance the phone's Augmented Reality capabilities. Here's all you need to know.
Recently posted . 990 views . 1 min read
Trending News & Articles
Rolls Royce cars are extremely luxurious. While there are many expensive pieces of equipment in Rolls Royce cars, their most relaxing feature is the silence that ...
Recently posted . 2K views . 2 min read
Tata Harrier’s three-row seat version in works, details out
Recently posted . 2K views . 0 min read
The following list of India's Top 5 Mobile Charger manufacture Brand 2019
Recently posted . 2K views . 0 min read
XUV300 is the latest entrant in the compact SUV segment.
Recently posted . 2K views . 0 min read
More in Electronics & Gadgets
WhatsApp has said that it will not work on `Nokia S40` after December 2018.
Recently posted. 762 views . 0 min read
Microsoft Corp.'s first-quarter sales and earnings topped analysts'estimates, buoyed by means of developing call for for cloud-primarily based software and ...
Recently posted. 681 views . 43 min read
At the launch of the OnePlus 6, OnePlus curiously and cautiously had stayed away from using the word “Dash Charge” to address its unique fast charging i...
Recently posted. 859 views . 2 min read
Recently posted . 1K views
Recently posted . 943 views
Recently posted . 1K views
Recently posted . 1K views
Recently posted . 1K views . 78 min read
Recently posted . 1K views . 39 min read
NEW DELHI: Stuck in a bad signal area, or getting the dreaded sign that there is no mobile network at all? Soon you'll be able to latch on to...
Recently posted. 730 views . 2 min read
You know how sometimes when you are using Instagram and have to text someone but have to switch to WhatsApp because Insta chat is kind of limiting? Yeah, well, Fa...
Recently posted. 1K views . 1 min read