Turbidity Sensor Calibration Procedure at Instronline
Electronics & Gadgets  
instronline

The turbidity method is based upon a comparison of intensity of light scattered by a sample under defined conditions with the intensity of light scattered by a standard reference suspension. A turbidimeter is a nephelometer with a visible light source for illuminating the sample and one or more photo-electric detectors placed ninety degrees to the path of the light source. Note: the below calibration procedure is for a turbidimeter which the sample is placed into a cuvette.

 

Some instruments will only accept one standard. For those instruments, the second, third, etc., standards will serve as check points.
 
Calibration Procedure:
 
1. Allow the calibration standards to equilibrate at the ambient temperature. The use of commercially available polymer primary standards (AMCO-AEPA-l) is preferred; however, the standards can be prepared using Formazin (read the warning on the label before use) according to the EPA analytical Method 180.1. Other standards may be used if they can be shown that they are equivalent to the previously mentioned standards.
 
2. If the standard cuvette is not sealed, rinse a cuvette with deionized water. Shake the cuvette to remove as much water as possible. Do not wipe dry the inside of the cuvette because lint from the wipe may remain in the cuvette. Add the standard to the cuvette.
 
3. Before performing the calibration procedure, make sure the cuvettes are not scratched and the outside surfaces are dry and free from fingerprints and dust. If the cuvette is scratched or dirty, discard or clean the cuvette respectively. Note: some . manufacturers require the cuvette to be orientated in the instrument in a particular direction for accurate reading.
 
 

 

4. Select a low value standard such as a zero or 0.02 NTU and calibrate according to manufacturer’s instructions. Note: a zero standard (approximately 0 NTU) can be prepared by passing distilled water through a 0.45 micron pore size membrane filter.
 
5. Select a high standard and calibrate according to manufacturer’s instructions or verify the calibration if instrument will not accept a second standard. In verifying, the instrument should read the standard value to within the specifications of the instrument. If the instrument has range of scales, check each range that will be used during the sampling event with a standard that falls within that range.
 
6. Record the calibration information on calibration log sheet.

 
 


 
 


 
More in Electronics & Gadgets
Delete These 13 Gaming Apps From Your Android Phone, As It's Malware A...

Malware is an ever-present threat online, sort of like the common cold. It's all around, and you can never really predict it's arrival until too late. All...

Recently posted . 4 views

New WhatsApp mode coming soon, will help you during your critically im...

WhatsApp Vacation Mode will enable users to completely mute their group notifications, meaning no pop-ups would appear on the phone in either the locked screen or...

Recently posted . 4 views

You will be banned on WhatsApp if you did these things; why you should...

you thought for once, WhatsApp does not care of what content you share with other people, then you are wrong.   

Recently posted . 28 views

Instagram will remove fake likes and follows

Instagram wants to fight fake likes, follows, and comments by more actively removing them. Any fake likes and follows from accounts that use third-party services to...

Recently posted . 9 views

 
 
 

Prashnavali

Thought of the day

"I've learned that people will forget what you said or what you did, but they'll never forget how you made them feel."
Maya Angelou