Differential Pressure Sensors Calibration Procedure

Differential pressure sensors are common in the process industry and cover a variety of applications. To understand what a differential pressure sensor is, it becomes important to put it in contrast to other pressure measurement types. The most common types of pressure measurement are absolute, gauge and differential.
 

 
 

Gauge Pressure: Gauge pressure is the pressure difference in reference to barometric (or atmospheric) pressure as showing in figure 1. This is the most common pressure measurement type in industry today.
 
Absolute Pressure: Absolute pressure is when zero pressure is referenced to absolute vacuum as shown in figure 1. This is done by pulling a very hard vacuum, achieving as close to absolute zero as possible, and then referencing the zero of the sensor to that vacuum point. Often absolute sensors utilize a gauge sensor and a barometric sensor and calculate the absolute pressure by subtracting the barometric pressure from the gauge pressure.
 
Differential Pressure: Differential pressure (DP) can be independent of the atmospheric and absolute pressures. It is the pressure difference between two applied pressures and as shown in figure 1. These sensors are very useful in determining the pressure difference between two places or systems and are often used in flow calculation, filtering, fluid level, density, and viscosity.
 
So now that we’ve reviewed the different pressure types and we know what differential pressure is and how it compares to other pressure measurement types. Now, we can consider how we calibrate a DP sensor and some of the challenges associated with calibration of DP sensors. First, let’s start with the challenges.
 
Common Challenges in Calibrating DP Sensors
 
Producing a stable, controlled pressure – to have a meaningful measurement for calibration we must be able to have stable pressure generation from a pressure source, such as a pump or a controller. DP sensors can be very sensitive, so a solution that will produce and hold a stable pressure is very important. Also, the pump or controller needs to have sufficient resolution to be able to exactly generate the desired pressure points. Producing a stable, controlled pressure with high resolution is often a challenge because many pump solutions rely on check valves, or non-returning valves, within the pump as shown in Figure 2 : DP gauge and pump stability. These check valves are prone to leaks over time and use and are often the source of frustration when trying to hold highly stable pressures for a DP sensors calibration.
 
Temperature effects – Possibly the largest challenge to calibrating DP sensors has to do with the impact of the environmental temperature on the DP sensor and the calibration standards. Because many DP sensors are measuring very low full scale (FS) pressures, a small change in temperature can amount to a very noticeable change in pressure. This change in temperature often equates to constant instability in both the sensor being tested and the calibration standard (both reference gauge and pump).
 
Changing atmospheric pressure – Several DP sensor manufacturers recommend the calibration be performed with the reference port (or low port) be open to atmosphere. The challenge with this requirement is that throughout a calibration, the atmospheric pressure is constantly changing which influences the stability and repeatability of the calibration results.
 
Methods of Calibration
Example 1 – Using an pressure pump, DP reference gauge with the DUT’s reference port open to atmosphere
 
Required Equipment:
 
 Low pressure calibration pump
 
 Device under test
 
 Reference DP Gauge
 
 Lines and fittings to connect from the gauges to the pump
 
Connection 
 
 Both the high ports of each gauge are connected into the calibration pump
 
 The reference or low ports of each gauge are left open to atmosphere
 
 Ensure the DUT is in the proper orientation (typically vertical or horizontal)
 
Procedure
 
 Depending on the DUT, you may need to exercise the gauge multiple times to its full scale.
 
 Ensure the vent valve is open and zero both the reference gauge and the DUT (assuming the DUT is a digital gauge that requires regular zeroing).
 
 Close the vent valve and proceed to the next calibration points and record the data when the measurement is stable.
 
 Increase the Pressure by using pump and note down the readings in DUT & Reference gauge.
 
 Typically, 3-5 calibration points are taken both upward then downward so as to determine hysteresis.
 
Pros: This method is inexpensive and the set up is easy.
 
Cons: You’ll need to account for barometric pressure and temperature changes throughout the test. Depending on the environmental conditions this can produce very unstable measurements. This is the least accurate method for calibration of DP sensors.
 
Example 2 – Using an Low pressure calibration pump, DP reference gauge with the DUT’s reference ports connected together
 
Required Equipment:
 
 Low pressure calibration pump
 
 Device under test
 
 Reference DP Gauge
 
 Lines and fittings to connect from the gauges to the pump and the gauges together
 
Connection 
 
 Both high ports of each gauge are connected into the calibration pump.
 
 The reference or low ports of each gauge are connected together.
 
 Ensure the DUT is in the proper orientation (typically vertical or horizontal).
 
Note: In this method pressure is generated on both the high and low pressure lines and the DP is measured by the reference gauge. Depending on the DP range required the best solution to reach the full scale of the DUT.
 
Differential Pressure Sensor Calibration
 
Procedure
 
 Depending on the DUT, you may need to exercise the gauge multiple times to its full scale
 
 Recording the zero point may vary depending on the type of DUT. If the DUT is a digital gauge, then keep the reference gauge and the DUT reference ports connected together and zero both gauges. If the DUT is an analog gauge that doesn’t require a regular zero, then disconnect both reference ports and leave them open to atmosphere to zero the gauges. After recording the zero point connect both the reference ports together and proceed through the calibration.
 
 Close the vent valve and proceed to the next calibration points and record the data when the measurement is stable.
 
 Increase the Pressure by using pump and note down the readings in DUT & Reference gauge.
 
 Typically, 3-5 calibration points are taken both upward then downward so as to determine hysteresis
 
Pros: This method is inexpensive and better accounts for atmospheric pressure changes throughout the test. The stability at each point is improved from the first example.
 
Cons: The set up is more complicated than the first example and temperature effects can potentially have a larger impact than the first example because we have a sealed system with the low (reference) lines being connected.
 
Example 3 – Using the automated calibration Equipment
 
Required Equipment:
 
 Automatic Calibrator
 
 Device under test
 
 Lines and fittings to connect the DP gauge to the Calibrator
 
Connection 
 
 Connect the high port of the DP gauge to the OUTLET port of the Automatic Calibrator Equipment
 
 Connect the low port of the DP gauge to the REF port of the Automatic Calibrator Equipment.
 
 Ensure the DUT is in the proper orientation (typically vertical or horizontal).
 
 

 
 

Calibrator with DUT
 
Procedure
 
 Depending on the DUT, you may need to exercise the gauge multiple times to its full scale.
 
 Program in a task and run an automated test which will automatically generate the pressure, stabilize the measurement, and allow for the DP gauge reading to be recorded.
 
 Typically, 3-5 calibration points are taken both upward then downward and the Calibrator will automatically calculate the hysteresis and display the test results with pass/fail criteria.
 
Pros: This method is fully- or semi-automated depending on the DUT. Measurements are controlled and stability is ensured by the Calibrator controller. The Calibrator is much less influenced by changes in temperature and barometric pressure than the previous examples. Results are automatically displayed and calculated. The Calibrator can calibrate pressure gauges and transmitters.

 
 
 
 
 

Related Topics

 
 
 

Trending News & Articles

 Article
'Worse than prison': A rare look inside China's detention camps to 'brainwash' Muslims

ALMATY: Hour upon hour, day upon day, Omir Bekali and other detainees in far western China's new indoctrination camps had to disavow the...

Recently posted . 202K views . 1 min read
 

 Article
What The Shape Of Your Belly Button Says About Your Health

If you have payed attention to the belly buttons of people on the beach or the members of your family, you have probably noticed that they have different shapes and...

Recently posted . 9K views . 2 min read
 

 Article
Top 10 Horrifying Acts of Chemical Warfare and Gas Attacks

In this age of terror, there might be nothing more terrifying than the thought of an attack carried out with chemical weapons. We’ve all heard the horrific ...

Recently posted . 4K views . 4 min read
 

 Article
Top 10 Best Gym Equipment Brands in India 2018

Body fitness is one thing that everyone wants to maintain irrespective of age. Going to the gym and doing some great exercise always helps to maintain your body fit...

Recently posted . 4K views . 2 min read
 

 
 

More in Electronics & Gadgets

 Article
Lakes and Labyrinths: NASA Reveals First Geological Map of Saturn’s Planet-sized Moon 'Titan'

The largest moon of Saturn is the only planetary body in the solar system, apart from Earth, that has stable liquid lakes on its surface and dunes that are made up ...

Recently posted. 684 views . 1 min read
 

 Article
NASA Pic Shows Us What Earth Would Look Like If All Oceans Dried And It's Terrifying

When we see images of Earth from space, we see a big blue planet, with some brown terrain. This is our home, the only planet we can survive on.

Recently posted. 801 views . 1 min read
 

 Article
Scientists Find The First-Ever Animal That Doesn't Need Oxygen to Survive

Some truths about the Universe and our experience in it seem immutable. The sky is up. Gravity sucks. Nothing can travel faster than light. Multicellular life nee...

Recently posted. 638 views . 1 min read
 

 Reviews
The Best 5 Hiking Backpacks in India – Reviews & Buying Guide



Recently posted . 2K views . 140 min read
 

 Article
Is Life difficult?

Life……i think the most easy expression to define the same is the BREATH, which is coming and going at its pace, without any visible effort from us.<...

Recently posted. 1K views . 1 min read
 

 Article
Pakistan needs to stop thinking of Kashmir as an unfinished business of Partition

Pakistan has repeatedly altered the status of the parts of Kashmir it controls, weakening its current protestations.

Recently posted. 699 views . 0 min read
 

 
 
 

   Prashnavali

  Thought of the Day

Thank you for helping me build my life.
Anonymous

Be the first one to comment on this story

Close
Post Comment
Shibu Chandran
2 hours ago

Serving political interests in another person's illness is the lowest form of human value. A 70+ y old lady has cancer.

November 28, 2016 05:00 IST
Shibu Chandran
2 hours ago

Serving political interests in another person's illness is the lowest form of human value. A 70+ y old lady has cancer.

November 28, 2016 05:00 IST
Shibu Chandran
2 hours ago

Serving political interests in another person's illness is the lowest form of human value. A 70+ y old lady has cancer.

November 28, 2016 05:00 IST
Shibu Chandran
2 hours ago

Serving political interests in another person's illness is the lowest form of human value. A 70+ y old lady has cancer.

November 28, 2016 05:00 IST


ads
Back To Top